Description

FEATURES
  • Multi-functions : WATT, VA, Whr, PF ( Power factor ),ACV, ACA, DCV, DCA, Hz, ohm.
  • True AC power( Watt ) & apparent power ( VA ) measurement.
  • True rms display for ACV, ACA.
  • 0.1 W resolution ( <1000 W ), high precision and high
    resolution in the low watt range, good performance ,
    for low power LED lamp watt measurement.
  • Supper large LCD, easy to read-out, display the Watt,Power factor, Voltage & Current value at the same time.
  • Accept different kinds current input signal as direct input,inductive clamp probe or CT ( current transformer ).
  • Auto range.
  • Built-in peak hold & data hold function.
  • Watt & VA measurement with Hi, low alarm setting capability
  • RS-232 output interface.
  • Exclusive custom exclusive design LSI circuit, provides high accuracy, reliability and durability.
  • Built-in over input indication.
  • Power supply by batteries or AC to DC adapter.
  • Built-in low battery indicator .
  • Durable bench type housing plastic case with carrying handle.
GENERAL SPECIFICATIONS
Display 93 mm x 52 mm large LCD ( Liquid Crystal Display) display.
Multi-display unit, show Volt, Ampere,
Watt, Power factor or Hz at same time.
Measurement WATT, VA, Whr,, Power factor, ACV, ACA,DCV, DCA, Hz, ohm.
Zero Adjustment Watt :
Watt : External adjustment by push button.
DCV, ACV, DCA, ACA :
Automatic adjustment.
Polarity Automatic switching, “-” indicates reverse polarity.
Current input mode Direct input, inductive clamp probe or CT.
Over input
Indication
Indication of “- – – – or “- – – –
Data Output RS232 serial interface.
Sampling Time W, VA, ACA, ACV, PF, Hz :
Approx. 1.5 Sec
DCV, DCA, OHM :
Approx. 1 Sec.
Operating Temp. 0 to 50 °C ( 32 to 122 °F ).
Operating
Humidity
Less than 80 % R.H..
Power Supply Battery power :
DC 9V, 1.5 V AA (UM-3) battery x 6 PCs.
AC power :
AC to DC 9V adapter ( 500 mA ), optional.
Power
Consumption
Battery power :
Approx. DC 50 mA
Dimension 28 x 21 x 9 cm ( 11.0 x 8.3 x 3.5 inch ).
Weight Approx. 1.6 Kg ( 3.52 LB ).
Standard
Accessories
Test lead ( red & black )……………….. 1 pair.
Instruction Manual………………………. 1 PC.
OHMS
Range Resolution Accuracy
9,999 ohm 1 ohm ± ( 1 % + 1d )
19.99 K ohm 10 ohm
Auto range.
Overload protection ” Max. AC/DC 300 V.
ELECTRICAL SPECIFICATIONS ( 23± 5 °C )
Watt (AC, true power),
current mode from direct input
Range Resolution Accuracy
6,000 Watt 0.1 W (< 1000W ) ± ( 1.5% + 5 d )
0.1 W (< 1000W )
Accuracy are specified under the following
conditions :
a) AC input current is 0.01 ACA & 10 ACA.
b) AC input voltage is within 110 V± 15 % and 220V± 15%.
c) ACA, ACV input signal is sine wave, 50/60 Hz.
d) Power factor 0.5.
ACA, ACV frequency response is from 40 to 400 Hz.
Max. volt & current input signal value :
Volt input : Max. AC 600V, Current input : Max. AC 10 A
Watt (AC, true power),
current input cooperate with inductive probe or CT
Range Resolution
0.1 to 999,9 Watt 0.1 Watt
9,999 Watt 1 Watt
99.99 KW 0.01 KW
999.9 KW 0.1 kW
Accuracy will be same as the abve ” Direct Current
Input Mode ” but plus the accuray value of Current Transformer ( CT ) or the acccuracy value of Inductive Current Probe.
Input current should obey :
Inductive Probe – 20 ACA.
CT 100/5 A – 8 ACA
CT 1000/5 A – 80 ACA.
VA ( AC, Apparent Power ),
current mode from direct input
Range Resolution Accuracy
99.99 VA 0.01 VA ± ( 2 % + 2d )
999.9 VA 0.1 VA
9,999 VA 1 VA
Accuracy are specified under the following
conditions :
a) AC input current is 0.4 ACA & 10 ACA.
b) AC input voltage is within 110 V 15 % and 220V 15%.
c) ACA, ACV input signal is sine wave, 50/60 Hz.
ACA, ACV frequency response is from 40 to 400 Hz.
POWER FACTOR
current mode from direct input only
Range Resolution Accuracy
0.01 to 1.00 0.01 ± ( 1.5% + 2 d )
Accuracy are specified under the following
conditions :
a) AC input current is 0.01 ACA & 10 ACA.
b) AC input voltage is within 110 V 15 % and 220V 15%.
c) ACA, ACV input signal is sine wave, 50/60 Hz.
Max. volt & current input value :
Volt input : AC 600V, Current input : AC 10A
Hz
Range Resolution Accuracy
10.0 Hz to 99.9 Hz. 0.1 Hz ± ( 1 % + 1d )
100 Hz to 999 Hz. 1 Hz
Auto range.
Frequency signal input voltage level should > 6V & 600 V.
AC VOLTAGE ( true rms ),DC VOLTAGE
Range Resolution Accuracy
0.1 V to 299.9 V 0.1 V DCV :
± ( 1 % + 1d )
ACV ( 1( 10 V ) :
± ( 1 % + 7d )
ACV ( 1( 11 V to 100 V ) :
± ( 1 % + 5d )
ACV((Others:))
± ( 1 % + 1d )
300 V to 600 V 1 V
AC CURRENT ( true rms ),DC CURRENT
current mode from direct input
Range Resolution Accuracy
ACA,0.05 A to 1.999 A,2.00 A to 10.00 A 1 mA
10 mA
± (1%+3d)
DCA,0.01 A to 10.00 A 10 mA ± (1%+1d)
Max. input current : AC 10 A, DC 10 A.
ACA accuracy is test under input signal is sine wave,
50/60 Hz.
ACA frequency response is from 40 to 400 Hz.
ACA is true rms.
AC CURRENT ( true rms ),DC CURRENT
current mode from inductive probe
Range Resolution
ACA < 2 A
2 A to 1000 A
1000 A
0.01 A
0.1 A
1 A
DCA 1000 A 1A
Accuracy : Meter voltage range accuracy plus
Inductive Probe’s accuracy.
ACA is true rms.
AC CURRENT
current mode from CT ( current transformer )
Range Resolution
CT 100/5A, 0.1 – 200.0 A 0.01 A, < 20 A
0.1 A, ≧ 20 A
CT 1000/5A, 1 – 2000 A 0.1 A, , < 20 A
1 A, ≧ 200 A
Accuracy : Meter current range accuracy plus
CT ( current transformer ) accuracy.
ACA is true rms.
Watt Hour
current from direct input
Range Resolution
0.001 Whr to 9.999 Whr 0.001 Whr
10.00 Whr to 99.99 Whr 0.01 Whr
100.0 Whr to 999.9 Whr 0.1 Whr
1000 Whr to 9999 Whr 1 Whr
10 K Whr to 99.99 K Whr 10 Whr
100 K Whr to 999.9 K Whr 100 Whr
1000 K Whr to 9999 K Whr 1 K Whr
When Watt Hour value over 9999 K Whr, the Display value will reset to 0000 K Whr, then count up again.
Accuracy & other specification requirement same as ” Watt ” range exactly